Cybird XD : 플라스마 에뮬레이션 LED 경화등

Cybird XD : 플라스마 에뮬레이션 LED 경화등

SKU(재고 관리 코드):SKU: CXD-9000N

  • Polymerize light-cured materials in 1.5 sec. (2mm)
  • 2 Curing Modes:
    • High Power = 1,500 mW/cm^2 (5, 10, 15, 20 sec.)
    • Plasma Mode = 2,700 mW/cm^2 (1.5, 2, 2.5, 3 sec.)
  • Designed for super fast light curing and orthodontic bonding speed to reduce chair time
전체 세부 정보 보기

FAQ

Charger has a burning smell

When different adapter is used, since the voltages are different, it will burn your unit. It is very crucial that you use our adapters for our products only. Please contact us immediately.

Battery pins / Charger pins are corroded and will not charge    

Pins corrode due to too much sterilization.  Any liquid left from sterilization will cause corrosion.

Make sure to have the unit dry after sterilization.

Cleaning Instructions

Cleaning and Disinfection Instructions must be followed after each use.

Suggested Surface Disinfectant:

  • CaviCide®(or equivalent quaternary/alcohol blend disinfectant)


Do NOT Use:

  • Glutaraldehyde
  • Denatured alcohol
  • Lysol®
  • Phenol or phenolic cleaner
  • Ammonia complex
  • Iodine complex solutions

Disassembly and Inspection:

  1. Remove elastomer rubber light shield from glass fiber-optic light guide.
  2. Remove light guide from handpiece by gently pulling it straight out. Examine light guide ends for damage or composite material adhering to the tip. Examine light shield for tears or distortion. Replace if any damage is noted on either component.

Handpiece and Charger Base

Cleaning:

  1. Initial cleaning of handpiece must begin immediately after use to prevent drying of soil and contaminants in and on the device
  2. All exterior surfaces of the handpiece and charger base may be wiped with CaviWipes® or a cloth soaked in surface disinfectant to remove gross soil. Do not allow cleaning solution to invade the interior of these components since this may adversely affect the electronics.

Disinfecting:

  1. All exterior surfaces of the handpiece or charger base must be wiped and wetted with CaviWipes® or a cloth soaked in surface disinfectant.
  2. Allow the CaviCide® surface disinfectant to reside of the surface for a minimum of 3 minutes. Do not allow the disinfectant to dry on the surface.

Drying:

  1. Use a clean dry cloth to dry the exterior surfaces. Do not allow fluids to accumulate in the charger base socket as this may adversely affect the electronics.


Light Guide (Fiber Optic Glass)

Cleaning:

  1. Initial cleaning of the fiber optic light guide must begin immediately after use to prevent drying of soil and contaminants in and on the device.
  2. All exterior surfaces may be wiped with CaviWipes® or a cloth soaked in surface disinfectant to remove gross soil. Use a soft brush to remove contaminants in the junction between the glass rod and metal ferrule if necessary.
  3. Clean thoroughly using an ultrasonic cleaner. Minimum cycle time 10 minutes.

Rinsing:

  1. Using clean tap water on a clean cloth, wipe away residual disinfectant

Sterilizing:
Package in an FDA cleared wrap prior to sterilization. Sterilization can be performed in gravity autoclave at 132° C / 270° F for 15 minutes with a 15-30 minute drying time.

Light Shield (Elastomer Rubber)

Cleaning:

  1. Initial cleaning of the light shield must begin immediately after use to prevent drying of soil and contaminants in and on the device.
  2. All exterior surfaces may be wiped with CaviWipes® or a cloth soaked in surface disinfectant to remove gross soil. Examine the light shield for damage and discard if any cuts, tears, or distortion is noted.
  3. Clean thoroughly using an ultrasonic cleaner. Minimum cycle time is 10 mins.
  4. The elastomer rubber light shield can be cleaned, sterilized and reused up to 5 times, discard after that.

Rinsing:

  1. Using clean tap water on a clean cloth, wipe away residual disinfectant.

Sterilizing:

Package in a FDA cleared wrap prior to sterilization. Sterilization can be performed in a gravity autoclave at 132°C / 270°F for 15 mins. with a 15-30 min. drying time.

Reassembly & Storage

  1. After all parts have been allowed to dry completely, gently insert the metal end of the glass fiber optic light guide into the circular opening of the handpiece. Make sure the light guide is completely inserted until it stops. This will position the light guide properly with the internal LED light source.
  2. Insert an elastomer rubber light shield over the end of the fiber optic light guide.
  3. Store the reassembled Cybird Curing light in a clean, dry location, preferably on the Cybird charger base so that the battery will be completely charged prior to the next use.